Recall Z-1058-2017 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Brilliance iCT

DateDec 23, 2016
StatusTerminated
ReasonDuring a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued.
Root causeEmployee error
ActionPhilips sent an Customer Information letter. A document showed an incorrect torque wrench used during the assembly process which could potentially apply 4 Newton-meters less torque than specified to the system rotor counterweight bolts. Out of an abundance of caution, Philips will have a Field Service Engineer ensure the bolts on your system are correctly torqued. If you need any further information or support concerning this issue, please contact your local Philips representative. For further information please call (440) 483-5777.
Quantity1-unit
DistributionProduct was not distributed in the US. Unit was distributed in Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060937BRILLIANCE VOLUME
Philips Medical Systems (Cleveland), Inc., Jun 5, 2006