Recall Z-0987-2017 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

Oncentra External Beam Oncentra Brachy Product Usage: Oncentra is radiation therapy planning software designed to analyze and plan radiation treatment is three dimensions for the purpose of treating patients with cancer.

DateDec 23, 2016
StatusTerminated
ReasonCross profile for Varian 60 degree wedge shows "horns."
Root causeSoftware design
ActionElekta sent an Urgent Important Safety Notification letter dated December 23, 2016 to affected customers. The notice informs users of the specific product and version numbers affected by the issue and any work arounds that can be used to avoid the issue. The letter included an acknowledgement form to be returned to Elekta.
Quantity433
DistributionWorldwide Distribution - US Nationwide in the states of AL, CA, DC, NE, NY, ND, OR, PA, RI, SC, TX, and the countries of Angola, Argentina, Armenia, Australia, Austria, Barbados, Belarus, Brazil, Bulgaria, Canada, China Columbia, Curacao, Czech Republic, Denmark, Finland, France, Germany, Greece, Ho
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121448ONCENTRA, ONCENTRA EXTERNAL BEAM, ONCENTRA BRACHY
Elekta Inc. D/B/A Nucletron Corporation, Jul 2, 2012