Recall Z-1062-2017 ยท IZI System, X-Ray, Angiographic

Philips Electronics North America Corporation

Velara Generator with an old Firmware version used with the Philips Integris BH5000

DateDec 9, 2016
StatusTerminated
ReasonSystem may lock up.
Root causeDevice Design
ActionConsignees were notified of the recall by letter, sent via certified mail on 12/9/16. Philips will upgrade the firmware of the Velara Generator, allowing the Velara Generator to recover in case of intermittent shortcuts. This service will be provided free of charge for all affected systems and will start effective January 2017. A Philips service representative will contact customers with affected devices to arrange for the service.
Quantity59 devices (this number represents all products listed in recall)
DistributionUS: AL, CA, DC, IL,GA, KS, MD, MO, NC, NY, OH, PA, RI, SC, TX, WA OUS: Brazil, Bulgaria, Costa Rica, France, Germany, Israel, Italy, Japan, Korea, Republic of, Mexico, Russian Federation, Singapore, Spain, Sweden, Switzerland Switzerland, Turkey, United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033737ALLURA XPER FD20
Philips Medical Systems North America, Inc., Dec 9, 2003
K041949ALLURA XPER FD10
Philips Medical Systems North America Co., Jul 30, 2004