Recall Z-1090-2017 ยท JQP Calculator/Data Processing Module, For Clinical Use

Roche Diagnostics Corporation

cobas p 612 pre-analytical system; Pre-analytical sample handling that includes de-capping, Aliquotting and sorting of samples for analysis.

DateJan 24, 2017 (initiated Dec 27, 2016)
StatusTerminated
Reason
Root causeSoftware design
ActionOn 12/27/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected Consignees via courier fax or email. The recall notifications included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The cobas p 612 pre-analytical system (63x) software is being revised to correct the aspiration process. We anticipate this new version (2.0.4) to be available during Quarter 1 2017. For now, discontinue using disposable non-filter tips for Aliquotting on your cobas p 612 pre-ana
Quantity5
DistributionTX and WA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.