Recall Z-1386-2017 ยท DSP System, Balloon, Intra-Aortic And Control

Maquet Datascope Corp - Cardiac Assist Division

CARDIOSAVE Hybrid IABP and CARDIOSAVE Rescue IABP affected part number is 0998-:XX-0800-XX Li-Ion Battery affected part number is 0146-00-0097 Product Usage: The CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons

DateDec 23, 2016
StatusTerminated
ReasonMaquet has received 3 confirmed complaints related to the Li-lon Battery used with the Maquet CARDIOSAVE Hybrid IABP and Maquet CARDIOSAVE Rescue IABP of the battery venting after being accidentally dropped.
Root causeDevice Design
ActionMaquet contacted their customers via letter on 12/23/16 and additional customers identified were notified on 1/10/17 and issued a recall letter/return response form. The letter identified the affected product problem and actions to be taken. Customers were instructed to complete and return the response form via fax to 1-973-629-1518 or email to cardiosavebattery@getinge.com
Quantity8000
DistributionWorldwide Distribution - US Nationwide
FDA recordOpen on fda.gov