Recall Z-1096-2017 ยท LNH System, Nuclear Magnetic Resonance Imaging

Hitachi Medical Systems America Inc

Hitachi Oasis MRI System - C-Spine Coil

DateJan 6, 2017
StatusTerminated
ReasonThe customer indicated that the coil balun was hot to the touch when removing the coil from the patient table after an exam.
Root causeComponent design/selection
ActionHitachi sent an Urgent : Device Correction Notice dated January 3, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm plans to visit each site and replace the cable balun and also confirm receipt of the notification letter. The firm will quarantine returned baluns in their warehouse until they are reworked. For further questions regarding this recall please call (330) 425-1313 ext. 3720.
Quantity180 units
DistributionUSA (nationwide) distribution. The product was distributed to the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NY, OH, OK, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. There are three (3) government acc
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072279MAGNETIC RESONANCE DIAGNOSTIC DEVICE
Hitachi Medical Systems America, Inc., Sep 21, 2007