Recall Z-1093-2017 ยท JSG Culture Media, Non-Selective And Non-Differential
Acumedia Manufacturers, Inc.
Tryptic Soy Agar, Acumedia PN 7100 500g, 2Kg, 10Kg, and 50Kg sizes
| Date | Apr 22, 2016 |
| Status | Terminated |
| Reason | light precipitate may form in prepared plates of Acumedia Tryptic Soy Agar (7100) Lot# 108166. While there is no change in growth performance of this medium, the formation of particulate matter may be misinterpreted as growth in poured plates thereby leading to inaccurate counts. |
| Root cause | Component design/selection |
| Action | Neogen sent a Recall Notification dated March 8, 2016, to all affected customers on April 22, 2016. via e-mail. The recall notification included a description of the reason for the recall, affected product, and offer for replacement. If any Tryptic Soy Agar from this lot remains in your inventory, and you would like to replace this lot, please contact the customer service team at Neogen Corporation by email at foodsafety@neogen.com or by telephone at (800)-234-5333. We would like to assure you that we are committed to providing products that meet our strict quality standards for your test |
| Quantity | 482 |
| Distribution | Worldwide Distribution - US Distribution to the states of : AR, CA, CT, FL, GA, ID, IL, IN, MD, ME, MI, NC, ND, NY, OH, OR, TX, TN, UT, VA, WI. and to the countries of : CANADA, UK, ISRAEL, MEXICO, AUSTRALIA, COLOMBIA, TAIWAN, BRAZIL, CHILE and INDONESIA. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.