ClearanceBrief
Recall Z-1097-2017 ยท NSX

Soft Computer Consultants, Inc.

SCC Soft Computer Softbank II software Product Usage: Supports single and multi-site transfusion services in healthcare facilities. Used by healthcare personal to document, query, and view the integrated information regarding patients and products. Quality control testing, test and transfusion history, transfusion management, inventory management, product distribution, and final disposition are a

DateOct 6, 2016
StatusTerminated
ReasonSoftware error. Potential for incorrect results
Root causeSoftware design (manufacturing process)
ActionConsignees were sent a recall notification, dated 10/06/2016, providing instructions for software correction. The firm indicates that the issue will be resolved with a mandatory hot fix. Hot fix will be released when its effectiveness is confirmed. Consignees are instructed to acknowledge receipt of the recall notification and grant permission for the recalling firm to load the hot fix when available. The firm can be contacted by calling Technical Support at 1-800-763-8522. Until the hot fix is available, consignees can still operate the software by exiting (Esc-Quit) the Test Verify option af
Quantity189 total
DistributionWorldwide Distribution - US Nationwide Distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WA, WI, WV, and the countries of Hawaii and Puerto Rico Canada & Jamaica
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.