ClearanceBrief
Recall Z-1499-2017 ยท ILQ System, Communication, Powered

Stanley Security Solutions Inc

Arial Wireless Water-Resistant Call Pendant The Arial Pendant tag is part of the Arial wireless emergency call management system. It enables residents in assisted living, skilled nursing or independent living to call staff with the press of a button.

DateJun 22, 2016
StatusTerminated
ReasonDevices were incorrectly programmed during manufacturing therefore depressing the pendant button may result in an alarm not sounding as intended.
Root causeSoftware design (manufacturing process)
ActionStanley Healthcare sent an Important Recall Notice dated June 22, 2016, to all affected customers. The notice instructed customers to discontinue use of the device and to contact the recalling firm to arrange for return of the device. The recalling firm also asked the customer to complete a response form indicating they read, understood and complied with the recall notice. Customers with questions were instructed to contact Arial Customer Service at 888-716-0077. For questions regarding this recall call 781-216-9035.
Quantity500 individual pendants
DistributionNationwide Distribution to AR, AZ, CA, FL, GA, IA, ID, IN, KS, MD, MN, MO, NC, NE, NJ, NY, OH, OR, PA, SD, UT, VA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.