Recall Z-1055-2017 ยท JZQ Plates And Equipment, Radial Immunodiffusion

The Binding Site Group, Ltd.

Digital RID Plate Reader and Software Product Code: AD400

DateJan 19, 2017 (initiated Oct 9, 2012)
StatusTerminated
ReasonIf a control ring is marked after reading, the software will not flag results that are out of the specified QC range.
Root causeSoftware Design Change
ActionOn 10/2012 notification letter was sent to customers requesting they take the following actions: 1. If have not already installed software version Vl.0.3 then please continue to work with version V1.0.2 until the above issues are resolved. 2. If have already installed software version V1.0.3, when processing results always mark the controlprior to reading the ring. 3. If are uncertain about the validity of any assays you have performed using software version Vl.0.3 please review the controlresults obtained and compare them to the specified control range for the RID assay and; " If the
DistributionU.S. distribution to the following; MA, GA, NY, IA, CA, IL, UT, NJ, NC, MD, MN. Foreign distribution to the following; UK, Germany, Spain, France.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.