ClearanceBrief
Recall Z-1073-2017 ยท CFN Method, Nephelometric, Immunoglobulins (G, A, M)

The Binding Site Group, Ltd.

MininephThl Human lgG4 Kit Product Code: ZK009.LR This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents

DateJan 23, 2017 (initiated Sep 11, 2014)
StatusTerminated
ReasonCustomer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
Root causeNonconforming Material/Component
ActionBinding Site sent an Important Information letter dated September 30, 2014, Customers were instructed to retain the notification for records and additional instructions with the firm stating they were updating the instructions for use. Customers with questions were advised to contact their Binding Site representative or technical.service@bindingsite.co.uk.
DistributionWorldwide Distribution - US including n FL, OH, TN, MN, NW, MI, CA. and Internationally to Canada, France, UK, Spain, Germany, Czech Republic, Belgium, Italy, Australia, Brazil, Chile, China, Costa Rica, Kuwait, New Zealand, Oman, Pakistan, Portugal, Sweded, Taiwan, and Turkey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982100IGG AND IGG SUBCLASS NEPHELOMETRIC DIAGNOSTIC TEST KITS
The Binding Site, Ltd., Dec 20, 1999