Recall Z-1034-2017 ยท EID Syringe, Restorative And Impression Material

Ultradent Products, Inc.

NanoTips 30g 20pk, 100pk PN 1240, 1426 Description: NanoTips are injection molded royal blue tips that contain stainless steel 30-gauge cannula to aid in the delivery of chemicals. The NanoTip base is made from a polypropylene copolymer. The stainless steel cannula is made from 304 SS, and is 30-gauge in diameter and 0.488 in length. The cannula is secured in the base with an epoxy adhesive. Indi

DateJan 17, 2017 (initiated Dec 14, 2016)
StatusTerminated
ReasonUltradent Products, Inc. announces a voluntary field action for the NanoTips 30g 20pk, 100pk because the cannula attached to the tip hub may become loose or detach from the blue base that connects to the syringe.
Root causeDevice Design
ActionUltradent Products, Inc. sent an Urgent Medical Device Correction letter dated December 14, 2016, to all affected consignees. The letter instructed consignees to discard remaining inventory and return the response form. For any questions, please call at 801-553-4858.
Quantity143 packs of 100 tips; 104 packs of 20 tips
DistributionWorldwide Distribution - US (Nationwide) - Internationally to Belgium, Switzerland, Germany, Italy, South Korea, Netherlands, and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.