Recall Z-1026-2017 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Fresenius Medical Care Renal Therapies Group, LLC

Fresenius 2008T Series Hemodialysis System

DateDec 21, 2016
StatusTerminated
ReasonWhen the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine
Root causeDevice Design
ActionFresenius issued Customer Notification via Certified Mail on 12/21/2016. Users provided with problem and recommendation, when adjusting the UF Rate, UF Goal, or Tx Time, is to use the numeric data entry keys instead of the up or down arrow keys. Change the values by using data entry keys, then save by pressing CONFIRM. Fresenius will make arrangements to install software on the hemodialysis machine. Confirmation of the noticed by the consignee will be required via a Reply Form.
Quantity65361 units (US) 249 units (OUS)
DistributionNationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis Machine
Fresenius Medical Care Renal Therapies Group, LLC, Dec 10, 2015