Recall Z-0973-2017 ยท OWB Interventional Fluoroscopic X-Ray System

GE OEC Medical Systems, Inc

OEC 9800. MDL Numbers: D222250, D141598 The OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.

DateJan 9, 2017 (initiated Dec 19, 2016)
StatusTerminated
ReasonGE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.
Root causeUse error
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated December 19, 2016, to all affected consignees. GE informed consignees that GE will correct all affected products. If consignnes have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative. For questions regarding this recall call 801-536-4952.
DistributionWorldwide Distribution - US including U.S. States: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PUERTO RICO, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Internationally to Canada, Mexico, an
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111551OEC(R) 9800, OEC(R) 9800 PLUS
Ge Oec Medical Systems, Inc., Oct 4, 2011
K132027OEC 9800 PLUS
Ge Oec Medical Systems, Inc., May 1, 2014