Recall Z-0960-2017 ยท HWC Screw, Fixation, Bone

Trilliant Surgical Ltd.

Tiger Screw Sterilization Tray Base Assembly Model 210-00-001 Revision N, O The product is a component of a system indicated for the fixation of fractures, non-unions, arthrodeses, and osteotomies of bones appropriate for the size of the device.

DateJan 6, 2017 (initiated Dec 8, 2016)
StatusTerminated
ReasonFour lots of Tiger Screw Sterilization Tray Base Assemblies were released to the field with a printing error where k-wire description markings indicate diameters of 0.45 and 0.35 instead of 0.045" and 0.035", respectively.
Root causeUnder Investigation by firm
ActionTrilliant Surgical Ltd. sent an Urgent Medical Device Recall letter dated December 8, 2016, to all affected consignees. The letter instructed the consignees to 1) identify any/all affected product; 2) document any/all identified product via a provided response form and return to Trilliant Surgical; 3) exchange product upon receipt of replacements; and 4) return any/all affected product as originally identified and reported on the return response form. Customers with questions were instructed to call 800-495-2919.
Quantity50 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K153338Tiger Cannulated Screw System
Trilliant Surgical, Ltd., Jan 14, 2016