Recall Z-0977-2017 ยท PPM General Purpose Reagent

BioMerieux SA

NucliSENS magnetic extraction reagents, For in vitro diagnostic use, Product Usage: NucliSENS miniMAG is intended to be used for the extraction of total nucleic acids from biological specimens.

DateJan 13, 2017 (initiated Nov 23, 2016)
StatusTerminated
ReasonA drift of the performance with the BK test (BK virus, worst case application) was observed for some silica batches of Magnetic Silica (MagSil). This is the same issue as in recall Z-2342-2016.
Root causeNonconforming Material/Component
ActionConsignees were notified via, URGENT PRODUCT SAFETY CORRECTION NOTICE on/about 11/23/2016. This is the same issue as in recall Z-2342-2016, notification letter sent on/about 7/8/2016.
Quantity3723 units in total
DistributionWorldwide distribution. US Nationwide; BioMerieux Subsidiaries in: Argentina, Austria, Australia, Brazil, Canada, Switzerland, Chile, Colombia, Czech republic, Germany, Spain, France United Kingdom, Greece, China, Hungary, India, Italy, Japan, South Korea, Netherlands, Poland, Portugal, Russia Fed