ClearanceBrief
Recall Z-1004-2017 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Brilliance iCT Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories

DateNov 25, 2016
StatusTerminated
ReasonFailure to correctly document the installation of four M12 Bolts into the system rotor.
Root causeEmployee error
ActionPhilips sent a Customer Information Letter to their customer. The letter identified the affected product, problem description and actions to be taken by customers. For further information or support concerning this issue, contact your local Philips representative or local Philips Healthcare office. For SUPPORT North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377).
Quantity1
DistributionThe one system was distributed in Canada. There were no government or US distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060937BRILLIANCE VOLUME