Recall Z-1781-2017 ยท DQO Catheter, Intravascular, Diagnostic

Boston Scientific Corporation

Boston Scientific Imager II Angiographic Catheter, M001316581 (5-pack outer package UPN), M001316580, single unit, inner package UPN)

DateMar 23, 2017 (initiated Dec 13, 2016)
StatusTerminated
ReasonOne lot of Imager II Angiographic Catheters is being recalled due to packaging mislabeling. The product may be labeled as Imager II Contra 2 curve catheters, when in fact the package contains Contra curve catheters.
Root causeEmployee error
ActionCustomers were sent an Urgent Medical Device Removal letter, dated 12/5/2016, via overnight mail delivery beginning 12/15/2016. The letter identified affected product, described the issue, and asked for product to be returned. Removal instructions were provided and a Reply Verification Tracking Form was asked to be returned. Questions can be directed to Nicole Pshon, Field Action Manager at 763-494-1133 or BSCFieldActionCenter@bsci.com
Quantity26 boxes of 5 (130 singles)
DistributionAZ, FL, MO, NJ, NY, OH, PA, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121694IMAGER II ANGIOGRAPHIC CATHETER
Boston Scientific Corp, Sep 14, 2012