Recall Z-0991-2017 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
ConforMIS, Inc.
ConFORMIS ITOTAL PS-IPOLYIMPLANT KIT-RIGHT Catalog Number:TPS-111-1111 The iTotal PS Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental
| Date | Jan 10, 2017 (initiated Oct 31, 2016) |
| Status | Terminated |
| Reason | iView report displayed femoral cut measurements in the incorrect unit of measure (inches vs. mm). |
| Root cause | Software Design Change |
| Action | ConForMIS intiated the recall by telephone on October 31, 2016, to the Sales Rep to request return of the kit. For questions regarding this recall call 781-345-9001. |
| Quantity | 1 unit |
| Distribution | Distribution to TX only |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K161668 | ConforMIS iTotal(R) Posterior Stabilized (PS) Knee Replacement System (iTotal PS) Conformis, Inc., Jul 15, 2016 |