Recall Z-1046-2017 ยท LLZ System, Image Processing, Radiological

Merge Healthcare, Inc.

Merge Cardio software. Product Usage: Merge Cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.

DateMar 28, 2016
StatusTerminated
ReasonIf a reader selects Multi-study review prior to the original study completing its loading process and then immediately returns to image review of the original study, it is possible that not all current images will be present for review, which could result in incorrect treatment/diagnosis.
Root causeSoftware design
ActionThe recalling firm issued a letter dated 3/28/2016 via email or via certified mail if they did not have an email address for the customer. A second letter dated 7/11/2016 was issued to nonresponders.
Quantity188 sites potentially have the effected versions
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051649VERICIS CARDIOVASCULAR IMAGE AND INFORMATION SYSTEM, MODEL 4.0
Camtronics Medical Systems, Aug 2, 2005