Recall Z-0967-2017 ยท KNX Collector, Urine, (And Accessories) For Indwelling Catheter

C.R. Bard, Inc.

SURESTEP Tray, BARDEX I.C. COMPLETE CARE, Urine Meter, Temperature STATLOCK Stabilization Device

DateJul 11, 2016
StatusTerminated
ReasonThe product contained a latex catheter instead of a silicone catheter.
Root causeLabeling mix-ups
ActionBard Medical sent an Urgent : Medical Device Product Recall Letter dated July 11, 2016, to all affected customers. The letter requested that they check inventory and quarantine product, and if the product had been further distributed, they need to conduct a sub-recall. The letter included a reply form to be returned. For further questions, please call (770) 784-6220.
Quantity2690 units
DistributionUS Distribution to the states of : CA, CT, FL, GA, IL, IN, KS, KY, MD, MA, MI, NE, NV, NJ, NM, NY, OR, SC, TN, TX, UT, VA, WA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.