Recall Z-1100-2017 ยท KHO Fluorometer, For Clinical Use

PerkinElmer Health Sciences, Inc.

AutoDELFIA Plate Processor , Fluorometer, for clinical use Product Number: 1235-5110 Product Usage: The Wallac 1235 AutoDELFIA automatic immunoassay system is designed to automatically perform assays using the DELFIA technology. DELFIA technology is based on the proven and widely used method of time resolved fluorometry

DateDec 21, 2016
StatusTerminated
ReasonAutoDELFIA System with Version 3.0 Software: Potential plate loading failures
Root causeSoftware design
ActionPerkinElmer issued letter on 12/21/16 to customers with AutoDELFIA Systems running version 3.0 Manager Software (revisons 1 and 2). The issue cannot be corrected by the customer therefore it is recommended the AutoDELFIA System be operated in the default English language variant until the AutoDELFIA Manager Software can be updated. The AutoDELFIA System with version 3.0 Manager Software performs as intended when the selected language variant is English. Customers are recommended to implement a software update to be deployed by PerkinElmer to resolve the issue. AutoDELFIA Manager Software versi
Quantity1 unit (US) 6 (OUS)
DistributionWorldwide Distribution - US Nationwide in the state of MN and countries of China, India, Philippines, France, South Korea, Vietnam, Taiwan, Morocco, Kazachstan, UK, Spain, and Israel,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K935047AUTO DELFIA
Wallac OY, May 23, 1994