Recall Z-0878-2017 ยท LLZ System, Image Processing, Radiological

Merge Healthcare, Inc.

Merge OrthoCase software. The firm name on the label is Merge Healthcare, Hartland, WI.

DateDec 21, 2016 (initiated Oct 8, 2015)
StatusTerminated
ReasonMeasurements in the software are changing after saving a plan with a measurement result, which may result in potential patient injury or delay in diagnosis or treatment.
Root causeSoftware design
ActionMerge issued a letter dated March 28, 2016, via email on April 4, 2016, or via certified mail if they did not have an email address for the customer. The letter notifies the customer of the issue, informs them of a workaround and that a fix is not yet available but they will be notified once one is available. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. T
Quantity30 sites potentially have the effected versions
DistributionDistribution was made to medical facilities in CA, CO, CT, MA, MD, MN, MT, NE, NV, NY NJ, PA, SC, WA, and WI. Foreign distribution was made to Canada, as well as other countries.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022881CEDARA I-SOFTVIEW
Cedara Software Corp., Oct 22, 2002