Recall Z-1018-2017 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

The Binding Site Group, Ltd.

Optilite Analyser, Product Code: IE700 Product Usage: The Thermo Scientific Indiko Clinical Chemistry Analyzer is a fully automated random access analyzer used to measure a variety of analytes that may be adaptable to the analyzer depending on the reagent used. The Indiko Glucose (H-K) test system, is intended for in vitro diagnostic use in the quantitative determination of the glucose concentrat

DateMay 13, 2015
StatusTerminated
ReasonPotential risk of the Optilite lid/cover falling suddenly and causing injury when not placed in the fully open position.
Root causeComponent design/selection
ActionBinding Site sent an Urgent Field Safety Notice letter dated May 13, 2015 to their customers The letter identified the affected product, problem and actions to be taken. Customers are instructed to complete and return the Eback Form that this notice has been received and the actions to be taken have been completed to technical.services@bindingsite.co.uk
Quantity129 units
DistributionWorldwide Distribution - US Nationwide in the states of: CA, VA, GA, PA, NJ, KY, NY, AR, NM, NC, AL, GA, MI, and the countries of Canada, China, Australia, Israel, Ireland, UK, Germany, Belgium, Italy, France, Czech republic, Spain, Portugal.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110035INDIKO
Thermofisher Scientific OY, Jun 28, 2011