Recall Z-0970-2017 ยท DYB Introducer, Catheter

BARD PERIPHERAL VASCULAR, INC

Halo One 5F Introducer Sheath 10 cm 5 pack Product Code: HAL510F Product Usage: The Halo One Thin-Walled Guiding Sheath is designed to perform as both a guiding sheath and introducer sheath. The Halo One Thin-Walled Guiding Sheath is indicated for use in peripheral arterial and venous procedures requiring percutaneous introduction of intravascular devices. The Halo One Thin-Walled Guiding Sheath

DateJan 18, 2017 (initiated Dec 2, 2016)
StatusTerminated
ReasonComplaints of sheath separation, kinking, and/or tip damage during use.
Root causeDevice Design
ActionBard sent an Urgent Medical Device Recall Notification letter dated December 2, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to find attached instructions detailing the steps they need to take regarding the product. Customers with questions were instructed to contact their Bard representative. Customers were provided the following instructions: 1. Records show that your facility has purchased product codes affected by this voluntary recall. Do not use or further distribute any affec
Quantity101
DistributionNationwide Distribution to NY, MO, NY, NY, FL, KS, MI, TX, NV, OH, ME, WA, LA, AZ, UT, NH. No foreign distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161183Halo One Thin-Walled Guiding Sheath
C.R. Bard, Inc., Jun 3, 2016