Recall Z-0899-2017 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer America Inc

AQURE System Software Version 2.2.0 Model #: 933-599 UDI: (01)05700699335999(10)2.2.0 Catalog number: 933-599 The AQURE system manages Radiometer blood gas and immunoassay analyzers and results from 3rd party devices for point of care testing placed throughout the hospital. It enables the user to track connected devices, monitor performance and availability, and to quickly resolve and document i

DateDec 22, 2016 (initiated Nov 23, 2016)
StatusTerminated
ReasonDesign error when displaying additional information in the patient view window; error may result in misreading a parameter and its value.
Root causeSoftware design
ActionRadiomet4er sent a letter dated November 30, 2016, to all affected customers via email with a request for customers to complete a return form. Customers/Users were advised to not use the "more" button to expand patient results when viewing them in the AQURE patient view. Customers/Users Radiometer Representative will contact them to schedule a visit of a remote session. During the visit or remote session the representative will run a script on the AQURE system, which serve to remove the misalighnment possibility. It is noted in the letter to customers that expanding the results by using
Quantity7 total (2 in US)
DistributionWorldwide Distribution - US including MI and NY and Internationally to England, Switzerland, and United Arab Emirates.
FDA recordOpen on fda.gov