Recall Z-0810-2017 ยท DYB Introducer, Catheter

Oscor, Inc.

OSCOR BREEZEWAY 10F S60.8CM D65.6CM FIRM TIP Catalog # AB1061FT Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.

DateOct 5, 2016
StatusTerminated
ReasonComplaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.
Root causeNonconforming Material/Component
ActionOscor sent two(2) Product Recall notifications: -HOSPITALS: Immediately check your inventory to confirm that you do or do not have units from these lot(s) in your possession. Identify and set aside any unit from the identified lot in a manner that ensures the affected product will not be used. Check all storage and usage. Review, complete, sign and return the enclosed Acknowledgement Form attached in the letter, directly to Oscor Inc. at the fax number or e-mail on the form, for immediate replacement (as applicable). Who to call: Please call Oscor Inc. Customer Service Department at 727-937-2
Quantity255
DistributionWorldwide Distribution - US including CA, NJ, MD, TX, FL, NC, MO, and Puerto Rico. Internationally to AUSTRIA, BOLIVIA, CANADA, CHILE, CYPRUS, NETHERLANS, EGYPT, GERMANY, HONG KONG, ISRAEL, ITALY, SWITZERLAND, and VIETNAM