ClearanceBrief
Recall Z-0808-2017 ยท DYB Introducer, Catheter

Oscor, Inc.

OSCOR ADELANTE BREEZWAY 10F STRAIGHT S79CM D84CM Catalog # AB101219 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.

DateOct 5, 2016
StatusTerminated
ReasonComplaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.
Root causeNonconforming Material/Component
ActionOscor sent two(2) Product Recall notifications: -HOSPITALS: Immediately check your inventory to confirm that you do or do not have units from these lot(s) in your possession. Identify and set aside any unit from the identified lot in a manner that ensures the affected product will not be used. Check all storage and usage. Review, complete, sign and return the enclosed Acknowledgement Form attached in the letter, directly to Oscor Inc. at the fax number or e-mail on the form, for immediate replacement (as applicable). Who to call: Please call Oscor Inc. Customer Service Department at 727-937-2
Quantity33
DistributionWorldwide Distribution - US including CA, NJ, MD, TX, FL, NC, MO, and Puerto Rico. Internationally to AUSTRIA, BOLIVIA, CANADA, CHILE, CYPRUS, NETHERLANS, EGYPT, GERMANY, HONG KONG, ISRAEL, ITALY, SWITZERLAND, and VIETNAM
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101497DELIVERY SHEATH, MODEL ADELANTE BREEZEWAY
Oscor, Inc., Oct 20, 2010
K122958DELIVERY SHEATH, MODEL ADELANTE BREEZEWAY
Oscor, Inc., Dec 13, 2012