Recall Z-0786-2017 ยท KNT Tubes, Gastrointestinal (And Accessories)

AbbVie Inc.

AbbVie PEG Kit 20 FR

DateNov 22, 2016
StatusTerminated
ReasonAbbVie is issuing an urgent field correction notification due to an increased number of reports of disconnections of the AbbVie 20FR PEG tube from the Y-connectors.
Root causeDevice Design
ActionThe firm sent an URGENT: MEDICAL DEVICE FIELD CORRECTION NOTIFICATION AbbVie PEG Kit 20 FR to affected customers describing the reason for the notification. The letter asks affected customers to forward this letter to all Movement Disorder Specialists and physicians (e.g. GIs, IRs, and surgeons) that place PEG- J Tubes within your facility. The letter describes actions to be taken for patients still using the AbbVie PEG 20FR Tube. The letter requests user facilities to continue to monitor the performance of the Tube. The firm states that there is no need to replace the AbbVie PEG 20FR Tub
Quantity541 units
DistributionU.S. consignees located throughout the country and Puerto Rico. Foreign Consignees: Worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142793AbbVie PEG
AbbVie , Inc., Jan 14, 2015