Recall Z-1008-2017 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Smith & Nephew, Inc.

REFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116

DateNov 16, 2016
StatusTerminated
ReasonThe firm received complaints of cracks in the weld on the head of the mallet. In the reported cases, some of the lead beads and particles escaped from the mallet into the surgical wound.
Root causeProcess control
ActionSmith and Nephew sent an Urgent - Product Recall 1 st Notification dated November 16, 2016, to all affected customers. Actions for Hospital Representatives: 1. Please inspect your inventory and locate any unused devices from the above listed product and batch numbers, and quarantine them immediately. 2. Complete the last two columns in the Inventory Return Certification Form on the following page, indicating the quantities that need to be returned. If you do not have product to return, please place an x in the column No Product to Return. Include your contact information in the spaces provi
Quantity4444 units
DistributionWorldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the countries of : Argentina, Austria, Australia, Belgium, Bahrain, Brazil, Canada, China, Colombia, Costa Rica, Dubai, Denmark, Germany, Finland, France, Great Britain, Greece, Hong Kong, Ireland, Israel, India, It
FDA recordOpen on fda.gov