Recall Z-0789-2017 ยท FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Smiths Medical ASD Inc.
Acuvance2 Safety Catheter, Product No. 1720-AI Product Usage: The Acuvance 2 Safety Catheter is a Peripheral Intravenous Catheter (PIVC). PIVCs are designed for single use venipuncture on all patient populations where access to the veins and arteries are required. The catheter has a Safety Needle Assembly (SAN) designed to reduce risk of accidental needle sticks to the operators. The lateral port
| Date | Dec 19, 2016 (initiated Nov 21, 2016) |
| Status | Terminated |
| Reason | Smiths Medical became aware that three (3) lots of Acuvance2 Safety Catheter, product number 1720-AI, contain devices with missing or incorrectly positioned side port valves. |
| Root cause | Process control |
| Action | Consignees were sent on 11/21/2016 a Smiths Medical "Urgent Medical Device Field Safety Notice: dated November 18, 2016. The letter was addressed to "Clinician who uses the Acuvance2 Safety Catheter". The letter described the Reason for Recall, Risk to Health, and Instructions to Customers. Requested consignees to complete and return the "Urgent Medical Device Field Safety Notice Response Form" within 10 days of receipt. For questions contact Stericycle via email at the smithsmedicalrecall@stericycle.com. |
| Quantity | 26,750 |
| Distribution | Foreign Distribution in the Italy, France, Switzerland, Germany, Sweden |