Recall Z-1225-2017 ยท JCM Control, Hemoglobin, Abnormal

Tosoh Bioscience, Inc.

Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing.

DateNov 7, 2016
StatusTerminated
ReasonThe device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of HbA1c, e.g. 9%, but not at the diagnostic medical decision point of 6.5%.
Root causeSoftware design
ActionRecall notification letters were distributed to customers on 11/7/16. The letters instructed customers to review the chormatography to determine whether or not a variant peak was detected. If so, restest the sample using an alternative method.
Quantity682 analyzers
DistributionWorldwide distribution. The products were shipped to the following states: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, AND WY. The recalled products were s
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071132G8 AUTOMATED HPLC ANALYZER, MODEL HLC-723 G8
Tosoh Bioscience, Inc., Sep 16, 2008
K131580AUTOMATED GYLCOHEMOGLOBIN ANALYZER HLC-723G8
Tosoh Bioscience, Inc., Jan 23, 2014