Recall Z-0668-2017 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

Monaco TRP System The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.

DateNov 17, 2016
StatusTerminated
ReasonContours are not handled correctly if the CT dataset contains slices with DICOM Z coordinates of xx.x50mm.
Root causeSoftware design
ActionElekta sent an Important Field Safety Notice 382-01-MON-007 was sent to all affected customers on November 17, 2016, to all affected customers. The notice informs users of the specific product version numbers affected by the issue, and any work arounds that can be used to avoid the issue. The notice included a reply form which is to be returned.
Quantity2148
DistributionWorldwide Distribution - US including AK, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MD, MA, MI, MN, MO, MT, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, SD, TX, UT, VA, WA, WI, Puerto Rico and Internationally to Algeria, Antiqua & Barbuda, Argentina, Australia, Austria, Bahrain, Belarus, Botswana, Brazil, Bul
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132971MONACO