Recall Z-0866-2017 ยท HIF Insufflator, Laparoscopic

Northgate Technologies, Inc.

Insufflator, laparoscopic The ConMed GS2000, Smith & Nephew 500, Novadaq 50L and Nebulae I insufflators are devices that provide CO2 gas distention of surgical cavities for diagnostic and/or operative endoscopy. The insufflator has multiple operating modes which can be used for the following procedures: General Laparoscopic, Pediatric Laparoscopic, Bariatric Laparoscopic and Minimally Invasive V

DateNov 21, 2016
StatusTerminated
ReasonDevice could cause a patient overpressure situation without any visual or auditory indication or warning
Root causeSoftware design
ActionNorthgate Technologies sent an Urgent Medical Device Recall Notification letter dated November 17, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customers. Customers were instructed to examine their records and ascertain the location of affected devices, cease distribution or use and quarantine those products appropriately. Ship all affected products in their possession to Northgate Technologies, Inc., 1591 Scottsdale Court, Elgin, IL 60123. To return product, complete the enclosed Recall Acknowledgement Form and fax it
Quantity49 devices
DistributionWorldwide Distribution - US including FL and MA. Internationally to Canada