Recall Z-0730-2017 ยท LLZ System, Image Processing, Radiological

Merge Healthcare, Inc.

Merge Cardio software. Merge Cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information from Camtronics and other vendors systems including images, hemodynamic studies and reports, measurements (via import from DICOM Structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (wavefo

DateOct 1, 2015
StatusTerminated
ReasonReporting feature times out after inactivity for more than an hour sending the user back to the study list, which causes all reporting data being entered to be lost.
Root causeSoftware design
ActionMERGE sent an Urgent Medical Device Recall letter dated March 28, 2016, to all affected customers. Non-responders were issued a second letter dated July 11, 2016, The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to reply using the enclosed form and the return addressed envelope and also to ensure that all users of the product were provided with the notification. If product was further distributed, customer were asked to identify their customers and notify them at once of the product recall. Customers with questions were a
Quantity116 sites potentially have the effected versions
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051649VERICIS CARDIOVASCULAR IMAGE AND INFORMATION SYSTEM, MODEL 4.0
Camtronics Medical Systems, Aug 2, 2005