ClearanceBrief
Recall Z-0724-2017 ยท KKX Drape, Surgical, Antimicrobial

Halyard Health, Inc

Halyard C-Section Drape with Clear Screen and Full Incise (Non-Sterile), Product Code 44965 NS; C-Section Drape with Clear Screen and Full Incise (Sterile), Product Code 44966 00; C-Section Drape with Clear Screen (Non-Sterile), Product Code 44967 NS; C-Section Drape with Clear Screen (Sterile), Product Code 44968 00; C-Section Drape with Pouch and Fenestration (Non-Sterile), Product Code 44977 NS

DateDec 8, 2016 (initiated Nov 9, 2016)
StatusTerminated
ReasonDrapes within the affected lots may include a manufacturing variation that prevents convenient removal of the coated release paper, which inhibits easy access to the surgical site.
Root causeNonconforming Material/Component
ActionAccounts were notified of the recall by letter on 11/9/2016. The consignees were instructed to quarantine and discontinue use of the recalled lots and to discard/destroy per the facility's standard disposal procedure for nonwoven materials. A response form was included which was to be sent back.
Quantity132 cases
DistributionUS Nationwide Distribution in the states of AL, FL, GA, IL, KS, MN, MO, NE, NJ, OH, PH, TN, VA, WI, WY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K862801SPUNGUARD PLUS SURGICAL DRAPE
Kimberly-Clark Corp., Aug 7, 1986