Recall Z-1203-2017 ยท DQK Computer, Diagnostic, Programmable

Merge Healthcare, Inc.

The Schiller PB 1000, also known as: Merge Hemo Patient Data Module (PDM), V2 model, a major component of the Merge Hemo software.

DateMay 31, 2015
StatusTerminated
ReasonNon-invasive blood pressure pumps (NIBP) voltage is not sufficient and causes the NIBP pumps to fail prior to reaching adequate initial pressure.
Root causeNonconforming Material/Component
ActionThe recalling firm issued letters via email dated 3/28/2016 and 7/11/2016, or via certified mail if no email address was available. A third letter dated 1/23/2017 was also issued..
Quantity341 sites potentially have the affected product versions
DistributionDistribution was nationwide to medical facilities. There was also military and government distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082421HEARTSUITE HEMODYNAMICS 7.10
Emageon, Inc., Dec 17, 2008