Recall Z-1082-2017 ยท PAD Chest Drainage Kit
Atrium Medical Corporation
Atrium Medical Ocean Blood Recovery Drains as follows: 2050-000 Ocean Blood Recovery drain with in-line connector and suction control stopcock 2050-300 Ocean Blood Recovery drain with suction control stopcock, no in-line connector 2052-000 Ocean Double Blood Recovery drain with in-line connector and suction control stopcock - 2 patient lines connected by one Y connector 2052-300 Ocean Double Blo
| Date | Nov 29, 2016 |
| Status | Terminated |
| Reason | Outer Packaging is not sterile |
| Root cause | Package design/selection |
| Action | Atrium Medical issued recall letter on 11/29/16 via two-day express mail delivery advising users: of the problem, updated instructions for use; the new Warning Label applied to the packaging; and requesting to post the warning label near all Atrium chest drain product inventory. A Field Notification Reply Form acknowledging receipt of the Field Notification was requested to be completed and returned. Questions contact Atrium/Maquet/Getinge chest drainage representative or Atrium Customer Service at 1-800-528-7486 (603-880-1433), Monday through Friday between 8:00 am and 5:00 pm (EST). |
| Quantity | 964,463 cases (US) 521,975 cases (OUS) in total |
| Distribution | Worldwide Distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, Chile, China, China, Columbia, Ecuador, Egypt, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Ireland, Italy, Kuwait, Lebanon, Libya, Mexico, Netherlands, New Zea |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.