Recall Z-1080-2017 ยท PAD Chest Drainage Kit
Atrium Medical Corporation
Atrium Medical Dual Chamber Ocean Drains as follows: 2020-000 Ocean dual chamber, in-line connectors and suction control stopcock (2 Patient Tubes) 2020-100 Ocean dual chamber with in-line connectors and no suction control stopcock (2 Patient Tubes) 2020-300 Ocean dual chamber with suction control stopcock, no in-line connector (2 Patient Tubes) Product Usage: To evacuate air and/or fluid from t
| Date | Nov 29, 2016 |
| Status | Terminated |
| Reason | Outer Packaging is not sterile |
| Root cause | Package design/selection |
| Action | Atrium Medical issued recall letter on 11/29/16 via two-day express mail delivery advising users: of the problem, updated instructions for use; the new Warning Label applied to the packaging; and requesting to post the warning label near all Atrium chest drain product inventory. A Field Notification Reply Form acknowledging receipt of the Field Notification was requested to be completed and returned. Questions contact Atrium/Maquet/Getinge chest drainage representative or Atrium Customer Service at 1-800-528-7486 (603-880-1433), Monday through Friday between 8:00 am and 5:00 pm (EST). |
| Quantity | 964,463 cases (US) 521,975 cases (OUS) in total |
| Distribution | Worldwide Distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, Chile, China, China, Columbia, Ecuador, Egypt, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Ireland, Italy, Kuwait, Lebanon, Libya, Mexico, Netherlands, New Zea |
| FDA record | Open on fda.gov |