Recall Z-0780-2017 ยท JQW Station, Pipetting And Diluting, For Clinical Use

PerkinElmer, Inc

PerkinElmer 20ul Clear RoboRack Tip, Pipette (Qty 960) Item Number: 6000677. Used with the: JANUS AUTOMATED WORKSTATION Product Codes: AJS4001, AJM4001, AJI4001, AJL4001, AJS8001, AJM8001, AJI8001, AJL8001, AJM4G01, AJI4G01, AJL4G01, AJM8G01, AJI8G01, AJL8G01, AJM4M01, AJI4M01, AJL4M01, AJM8M01, AJI8M01, AJL8M01 NGS EXPRESS Product Codes: AJS4NGS

DateOct 13, 2016
StatusTerminated
ReasonDefective tips have the potential to pipette/dispense volumes 50-100% lower than is specified for the tips.
Root causeProcess control
ActionPerkinElmer initiated the recall on Sept 1. 2016 by letter dated . Customers were told of the defect, potential for inaccurate measurements, and requested to inspect for the lots stop using and destroy. Replacement product to be provided. Accounts asked to compelte Response form. Questions call your PerkinElmer representative.
Quantity33312
DistributionIA, MD, NY, WA Foreign: France, Germany, The Netherlands, Hungary, Singapore
FDA recordOpen on fda.gov