ClearanceBrief
Recall Z-0682-2017 ยท DWC Controller, Temperature, Cardiopulmonary Bypass

Sorin Group USA, Inc.

STCKERT HeaterCooler System 3T; Item Number 160281, HeaterCooler 3T, 240V/60Hz; Item Number 160282, HeaterCooler 3T, 208V/60Hz; Item Number 160285, HeaterCooler 3T, 120V/ 60Hz The Heater Cooler 3T is used with a Stckert S3 heart lung machine and / or any other heart lung machine featuring a separate temperature control for extracorporeal perfusion of durations up to 6 hours

DateDec 1, 2016 (initiated Aug 1, 2014)
StatusTerminated
ReasonOn August 1, 2014 Sorin Group USA, Inc. informed consignees via letter to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU).
Root causeMaterial/Component Contamination
ActionSorin Group sent an Important Information letter dated August 1, 2014, to all affected consignees. The letter identified the product, the problem, and the action to be taken by then consignee. Consignees were instructed to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU). Consignees with questions were instructed to call 303-467-6527 or 303-467-6503. For questions regarding this recall call 281-228-7260.
Quantity1,618 units in the US; 129 units outside of US
DistributionWorldwide Distribution - US (Nationwide and Puerto Rico) and Internationally to Canada, Colombia, Ecuador, El Salvador, Germany, Honduras, Mexico, Saudi Arabia, Trinidad, and Tobago.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052601STOCKERT HEATER-COOLER SYSTEM 3T
Sorin Group Deutschland GmbH, Jun 6, 2006