Recall Z-0767-2017 ยท FOX Stand, Infusion

Firefly Medical, Inc.

IVEA 500A Intended for Med-Surg patient care.

DateDec 15, 2016 (initiated Nov 1, 2016)
StatusTerminated
ReasonThe dowel pin could potentially slip out of its press fit, resulting in the product potentially slumping to one side which could cause the patient to stumble.
Root causeProcess control
ActionFirm notified customers on **11/18/16** via certified "Urgent: Medical Device Recall (Field Correction) IVEA Model 500A" letter to correct product in the field. Customers were asked to thoroughly read the letter and to immediately follow the actions in the enclosed Product Correction Kit. The instructions include steps for cleaning the correction area and installing the permanent adhesive caps, as well as, completing and returning the product correction form to Firefly Medical. Customers may continue to use their units until the Product Correction actions are complete, as long as pins are n
Quantity239
DistributionWorldwide Distribution -- USA and Canada.
FDA recordOpen on fda.gov