ClearanceBrief
Recall Z-0604-2017 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SOMATOM Definition AS, Computed tomography system Product Usage: SOMATOM Definition Flash systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.

DateNov 17, 2016 (initiated Oct 19, 2016)
StatusTerminated
ReasonA problem can occur with the small light marker windows, used for the positioning laser and are integrated in the front cover of the SOMATOM Definition AS and Definition Flash systems, by either becoming loose or potentially dropping out. There is a risk that a person could touch rotating or electrical parts of the gantry if they were to reach through the window opening.
Root causeNonconforming Material/Component
ActionSiemens' sent a Customer Safety Advisory Notice letter dated October 19, 2016 to affected customers. The letter identified the affected product, problem, potential risks, how to avoid the potential risk related to this issue and actions to be taken.
Quantity762 systems distrtibuted in U.S.
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143400SOMATOM Definition AS/AS+
Siemens Medical Solutions USA, Inc., Apr 8, 2015