Recall Z-0646-2017 ยท GHH Fibrin Split Products

Lsi Medience Corporation

Pathfast D-Dimer, Reference number: PF1051-KUS

DateNov 21, 2016 (initiated Aug 16, 2016)
StatusTerminated
ReasonTwo complaints were received about decrease in sensitivity (control value and/or calibration value) of the product.
Root causeNonconforming Material/Component
ActionCustomers were notified by letter sent via email on/about August 1, 2016. They were instructed to stop using and destroy the affected kits. They were to perform a retrospective analysis of the patients tested on the batch and investigate further. A Destruction Certificate was attached to be completed and returned to Verbatim Americas, LLC. For further questions please call (704) 547-6517.
Quantity42 boxes (each containing 60 test cartridges/one kit contains 10 reagent cartridge trays and one tray has 6 cartridges).
DistributionUS Distribution to the states of : AL, FL, ID and MA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072288PATHFAST D-DIMER
Mitsubishi Kagaku Iatron, May 6, 2009