Recall Z-0369-2017 ยท KGI Densitometer, Bone

Medimaps Group

TBS iNsight Version v.3.0.1 Product Usage: TBS iNsight is a medical device software that is installed on bone densitometers for analysis of bone microarchitecture and osteoporosis management.

DateNov 5, 2016 (initiated Oct 3, 2016)
StatusTerminated
ReasonThe FRAX adjusted for TBS values are not correct when: The FRAX feature is activated in TBS iNsight; and TBS has been computed from a spine scan where some vertebrae were excluded.
Root causeUnder Investigation by firm
ActionMedimaps sent an Urgent Field Safety Notice letter dated October 3, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken.
Quantity15 units in US
DistributionWorldwide Distribution - US Nationwide - in the states of: IL, NC, CA, DE, NY, MD, WI, MN. And the following countries: France,Thailand, Croatia, Canada, Italy, Spain, Russian Federation, Ukraine, Lithuania, South Africa, Brazil, Lebanon, Saudi Arabia, Korea, China Portugalm Iraq, Syrian Arab Republ
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152299TBS iNsight
Medimaps Group SA, Apr 29, 2016