Recall Z-0747-2017 ยท JQW Station, Pipetting And Diluting, For Clinical Use

PerkinElmer, Inc

Perkin Elmer RoboRack 25ul, Clear, Non-Conductive, Filter Tips, Pre-sterilized Item Number: 6000689. Supplied for use with JANUS Automated Workstations: Product Number: AJS4001 AJM4001 AJI4001, AJL4001, AJS8001 , AJM8001, AJI8001, AJL8001, AJM4G01 , AJI4G01 , AJL4G01, AJM8G01, AJI8G01, AJL8G01, AJM4M01, AJI4M01 , AJL4M01, AJM8M01, AJI8M01, AJL8M01, AGS4NGS The JANUS Automated Workstation is an a

DateJun 24, 2016
StatusTerminated
ReasonFilter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inaccurate aspiration and dispensing
Root causeProcess control
ActionPerkinElmer initiated recall on June 24, 2016 to the Product Manager and was delivered to the customer by email facilitated through PerkinElmer Sales Team.Customers were requested to take the following actions until PerkinElmer's corrective measures are completed: Inspect customer inventory of the 6000689 25 Clear Filter Pipette Tip and identify if the affected lots are present. Complete and return the attached Response Form with the required information. Destroy all quantities of tips from the affected lots. Upon receipt of the return Response Form, a shipment of replacement tips will be sh
Quantity595 trays (571,200 tips)
DistributionWorldwide Distribution - US including CA, MA, NY, IA, OR , PA, WA and Internationally to Belgium, Canada, Finland, France, Germany, Japan, Netherlands, Singapore, and UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.