Recall Z-0430-2017 ยท DRS Transducer, Blood-Pressure, Extravascular

ICU Medical, Inc.

TRANSPAC IV MONITORING KIT NEONATAL, 24", DISPOSABLE TRANSDUCER, 30 ML SQUEEZE FLUSH (INFUSION PUMP), Item No. 42586-05 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a compatible pressure monitor. The SafeSet

DateNov 15, 2016 (initiated Sep 27, 2016)
StatusTerminated
ReasonICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
Root causeDevice Design
ActionICU Medical sent an Urgent Medical Device Recall Notification letter dated September 27, 2016, to all affected customers to inform them of the expanded recall of the monitoring kits with Transpac IV transducers. Customers were informed of updated communications from August 10, 2016, when ICU Medical sent a communication indicating a potential leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers. Based on continued evaluation, ICU Medical is expanding the notification to now include certain other monitoring kits that contain the Transpac IV transduc
Quantity800 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061573TRANSPAC III DISPOSABLE STRAIGHT PRESSURE TRANSDUCER AND KIT
Hospira, Inc., Sep 19, 2006