Recall Z-0411-2017 ยท MUJ System, Planning, Radiation Therapy Treatment

Mevion Medical Systems, Inc.

MEVION S250 Product Usage: Proton Radiation Therapy System

DateNov 15, 2016 (initiated Oct 20, 2016)
StatusTerminated
ReasonSoftware defect:2D projection of contours in Verity. The defect causes structures to appear off from their true position in the DRR and the radiograph with an error that increases with the distance of the structure from isocenter
Root causeSoftware design
ActionMevion issued an Important Safety Notice letter on October 14, 2016 to their customers. The letter identified the affected product, problem, containment, mitigation and actions taken. Customers were informed that Mevion will be updating software when it is released. For further Information, contact: srosenthal@mevion.com.
Quantity6 units
DistributionUS Nationwide Distribution in the states of: FL, MO, NJ, OH, OK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120676S-250 PROTON BEAM RADIATION THERAPY SYSTEM
Mevion Medical Systems, Jun 4, 2012