Recall Z-0589-2017 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)

Siemens Healthcare Diagnostics, Inc.

Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA Chemistry systems for the following methods: Albumin (ALB, ALB_c), Direct Bilirubin (DBIL_2), Total Bilirubin (TBIL_2), Calcium (CA, CA_c, CA_2, CA_2c), Cholesterol (CHOL), Creatinine (CREA, CREA_2, CRE_2c, ECRE_2), Glucose Hexakinase (GLUH), Glucose Oxidase (GLUO), Iron (IRON_2), Lithium (LITH), Magnesium (MG, MG_

DateOct 7, 2016
StatusTerminated
ReasonSiemens has confirmed that the DBIL_2 Assigned Calibrator Values (ACV) currently assigned to the affected lots may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for Direct Bilirubin measurement.
Root causeUnder Investigation by firm
ActionSiemens issued an Urgent Medical Device Correction in the US and an Urgent Field Safety Notice Outside the US beginning on October 6, 2016. Customers were notified of the issue and instructed to use the reassigned DBIL_2 calibrator value for Chemistry Calibrator Lot Numbers 680726, 680726A, 680726B, 680726C and 680726D. Customers were also requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter within 30 days by faxing to the Customer Care Center at 312-275-7795. For questions regarding this recall call 914-631-8000.
Quantity10,614 units
DistributionWorldwide Distribution - US (Nationwide) Internationally to Canada, Mexico, Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, Chile, Cyprus, Denmark, Egypt, Ecuador, Finland, France, Germany, Great Britain, Hong Kong, Hungary, Italy, Japan, Lithuania, Malaysia, Mexico, Netherlands, New Zeala
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050374ADVIA IMS LITHIUM ASSAY
Bayer Healthcare, LLC, Jun 15, 2005