Recall Z-0584-2017 ยท GXD Generator, Lesion, Radiofrequency

Stryker Instruments Div. of Stryker Corporation

RF MultiGen, 0406-900-000; Refurb RF MultiGen, 0406-900-000U

DateNov 16, 2016 (initiated Oct 11, 2016)
StatusTerminated
ReasonStryker Instruments is voluntarily recalling the Care Instructions/Instructions for Use packaged with the Radiofrequency (RF) MultiGen Generator, because the instructions for use incorrectly contained instructions supporting the sterilization of the MultiGen cables.
Root causeLabeling Change Control
ActionStryker Instruments initiated a voluntary recall of the RF MultiGen, and Refurb RF MultiGen because the instructions for use contained incorrect instructions on sterilizing the device. Stryker distributed customer notices via certified mail on 10/11/2016. Customer's were instructed to do the following: Immediately review this recall notification. 2. Enclosed was a revised version of the IVS Care Instructions. Please distribute the revised version to the appropriate person or department. 3. Remove and discard any previously distributed IVS Care Instructions or Instructions for Use. 4. Please
Quantity2,675
DistributionDomestic:AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY Foreign:France, Germany, Canada, UK, Lebanon, South Africa, Chile, Hong Kong, Poland, Argentin
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071482STRYKER RF MULTI-LESION GENERATOR, MODEL 0406-900-000
Synergetics, Inc., Dec 20, 2007